You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
Example Data File: 20200303 SPLP TCLP CMT ICP-MS
Batch Report
Instrument File Field Location
Universal Template File Field Location
Special Notes
Column G (Beginning with row 3)
Column A (Aliquot)
Column D (Beginning with row 3)
Column F (Analysis Date/Time)
Row 1 (Beginning with column H)
Column B (Analyte Identifier)
The number of columns may vary. Column H through the last column (Column EM in the example data file I have attached) will need to be parsed
Column H (Beginning with row 4)
Column C (Measured Value)
Convert blank cells, “N/A”, and “<0.00” to be imported as “0” Beginning with Column H of the instrument file, every other column will need to be parsed. For example, Column H values will need to be mapped starting in row 3, Column I will be skipped, Column J will be mapped, and Column K will be skipped, etc. If it is any easier for you, cell 2 in the column that needs parsed will read either “Conc. [ ppb ]” or “CPS” whereas cell 2 in the column that will be skipped will read either “Conc. RSD” or “CPS RSD” The number of columns may vary. Column H through the last column (Column EM in the example data file I have attached) will need to be parsed
The text was updated successfully, but these errors were encountered:
Parser Name: Agilent_7900_ICPMS
Example Data File: 20200303 SPLP TCLP CMT ICP-MS Batch Report