BK Precision 9130 Triple Output Power Supply
17-Apr-2015 K. Goetze
To run the BK9130 with EPICS (RS232):
====================================
RS232 Comms:
Baud: 9600
- The device can be set to 4800, 9600, 19.2K, or 38.4K. I tested at 9600.
Parity: No
- The device can be set to Odd, Even, or No Parity.
Stop bits: 1
Data bits: 8
Handshaking: No
inputEos: LF
outputEos: LF
Cable:
db9 DTE BK9130 db9*
------- ----------
2 2
3 3
5 5
* Notes Re. Serial connection:
1.) This device's db9 connector is NOT RS232. It is a TTL interface.
An adapter needs to be purchased from BK Precision that translates
the TTL to RS232. This adapter is port-powered. This means that,
for a low-power UART like the IP-Octal, +12VDC needs to be connected
on the RS232 side of the adapter to db9 pins 4 (+), and 5 (-).
Loading the database, for example:
# BK Precision 9130 Power Supply
dbLoadRecords("$(IP)/ipApp/Db/BK9130.db","P=xyz:,PS=1,PORT=serial2")
StreamDevice:
BK9130.proto must be in "STREAM_PROTOCOL_PATH"
Autosave:
BK9130_settings.req (can be used with "configMenu" to save instrument setups)
The arguments for the medm display are simply:
P=xyz:, PS=1