Thermo Scientific Neslab EX Setup Info
--------------------------------------
- StreamDevice RS232 support written by Kurt Goetze
- EPICS Files: NeslabEX.db, NeslabEX.proto, NeslabEX.adl, NeslabEX.req
- Neslab EX Digital Plus Controller Settings (set from front panel):
BAUD: 19200 (default)
SERL: 232 (default)
MODE: Computer (press this front panel button)
- In serial.cmd:
tyGSAsynInit("serial3", "UART_0", 2,19200,'N',1,8,'N',"","") /* Neslab EX Chiller*/
# Thermo Neslab Chiller streams support
dbLoadRecords("$(IP)/ipApp/Db/NeslabEX.db","P=xxx:,S=1,PORT=serial3")
- In auto_settings.req:
$(P)NL:1:T.DESC
$(P)NL:1:spT
$(P)NL:1:getT.SCAN
- RS-232 connection:
From the Manual, db9 pin #
2 - Tx
3 - Rx
5 - Gnd
- Note:
This is minimal support, based on the requirements of APS/sec.30.
Presently the functions supported are set/read temperature, readback
the programmed temperature, and warn and fail states for low H2O level.
More functionality may be added at a later date.